Chapter 8: troubleshooting – Labconco Bulk Powder Stations 3962622 User Manual

Page 40

Advertising
background image

Chapter 8: Troubleshooting

Product Service 1-800-522-7658

36

PROBLEM

CAUSE

CORRECTIVE ACTION

Lights do not work Lamp not installed properly.

Inspect lamp installation.

Lamp wiring disconnected.

Inspect lamp wiring.

Defective lamp.

Replace lamp.

Light switch is defective.

Replace light switch.

Defective electronic ballast.

Replace ballast.

Airflow monitor

malfunction

No power.

No lights.

No display.

Power supply is not plugged into proper voltage; plug in power

supply. Verify that airflow monitor interface cables are

connected. Check fuses on FilterMate or accessory light.

No audible alarm.

Alarm has been temporarily silenced using “test/reset” or “enter”

buttons.

Wrong alarm set point.

Airflow monitor was not properly adjusted. Repeat calibration

steps outlined in the User’s Manual.

Constant audible alarm.

Check airflow and calibration of airflow monitor.

Continuous alarm.

Check the face velocity of the enclosure as the airflow of the

system may have changed. If face velocity is correct, calibrate

the airflow monitor as outlined in the User’s Manual.

Monitor alarms; air way to

airflow monitor sensor is

blocked by insects, dust or

debris.

Lightly clean the airway with clean air. Be careful not to touch

sensitive electrical components.

Audible disable will not stay

ON.

An alarm condition must be continuously present before the

audible alarm can be silenced. If flow conditions fluctuate near

the alarm set point, the airflow monitor will automatically reset it.

Action should be taken to bring the enclosure airflow into proper

operating parameters or adjust the alarm set point lower.

Lift won’t operate

or malfunctions

No power.

Lift motor is not connected to a proper outlet. Connect the plug

into the four-plex electrical box located under work surface.

Intermittent operation.

Allow lift system’s thermal override to cool for 10 minutes then

adjust height.

Photoelectric eye

malfunctions.

Either adjust photoelectric eye sensitivity or replace

photoelectrical eye if it malfunctions. See replacement parts.

Hydraulic lines pinched

Replace entire lift system. See replacement parts.