Location and air current requirements, Background on electrostatics or static electricity – Labconco Bulk Powder Stations 3962622 User Manual

Page 9

Advertising
background image

Chapter 2: Prerequisites

Product Service 1-800-522-7658

5

To correct or ensure against electrostatic issues, the following additional measures
may be prescribed to improve weighing operations.

Maintain a humidity level between 45 and 60%. The ability to sustain this

humidity range can be challenging depending upon the regional climate and
HVAC system.

Ionizers in various forms (guns, bars and blowers) are effective ways to flood

an area with ions and essentially “neutralize” electrostatic electricity.

Background on Electrostatics or Static Electricity

Electrostatic charges on a surface such as the wall of a balance enclosure are not
created by moving air. Gases do not cause the charge. Impurities within the air
impinging upon surfaces dictate the polarity and magnitude of the charge. The
process, triboelectrification, occurs when the friction of the dust particles contact
the surface and electrons move across the interface.

The ability of a material to become polarized is a property known as permittivity.
On highly insulative materials like acrylic, ions or charged molecules are strongly
bound to the surface by polarization forces. The higher the force, the higher is the
permittivity value of the material. It is suggested that the use of high permittivity
materials, such as plastic, be avoided.

Since static electricity is a surface phenomenon, materials can also be classified by
their surface resistivity measured in ohms per square. The table below lists the
surface resistivity of various classes of material.

Surface Resistively Table

Material

Surface Resistivity

Example

Conductive

0

10

5

per square

Skin, Metals

Static dissipative

10

5

10

9

per square

Glass

Antistatic

10

9

10

12

per square

Polyethylene bag

Insulative

10

12

per square

Acrylic, Packing foam, Styrofoam

Location and Air Current Requirements

The XPert Bulk Powder Enclosure has been designed to contain hazardous
powders by negating typical cross drafts and movements within the opening. Air
movement does not affect most modern balances with draft shields. However, as a
precautionary measure of safety and a higher level of quality management, it is
recommended that the enclosure be placed in such an area to avoid:

High traffic areas where walking might cause an air disturbance or be a

nuisance to balance readings.

Overhead or wall HVAC diffusers, fans, radiators or other lab equipment

producing air currents.

Next to doorways or windows that may be opened.

Do not position the unit so that it is difficult to operate the main
disconnect device.

Ne placez pas l'appareil de sorte qu'il est difficile de faire
fonctionner le dispositif principal de déconnexion.

!

Advertising