Chapter 6: Maintaining Your Enclosure

Product Service 1-800-522-7658

26

Programming the Guardian 1000 Digital Alarm

Set the desired face velocity of the enclosure with the sash in the down position
using an average face velocity of readings from a thermal anemometer.

a.

2 ft enclosure – 2 readings

b.

3 ft enclosure – 3 readings

c.

4 ft enclosure – 4 readings


Configure Alarm

To successfully calibrate the airflow monitor, it will be necessary to change the
face velocity by adjusting the airflow exhaust volume. The exhaust volume can be
adjusted using the speed control located behind the front panel on the XPert
Filtered System or on the back of the FilterMate. For the XPert Filtered Station,
the exhaust volume is adjusted through a damper in the building exhaust system.

The following inflow face velocity speeds are recommended to successfully
calibrate. To begin, determine what the low airflow, low warning airflow, normal
airflow, and high airflow shall be.

Low Air Alarm

Set Point

(user defined)

Enclosure

Operating

Inflow Speed

Low Calibration

Set Point

(user defined)

High Calibration

Set Point

(user defined)

40 – 50 fpm

60 fpm

40 – 60 fpm

100 – 120 fpm

60 – 70 fpm

80 fpm

50 – 90 fpm

100 – 150 fpm

80 - 90 fpm

100 fpm

50 – 110 fpm

100 – 170 fpm


NOTE: ENTER button stores information

+/- Buttons allow for scrolling.

1.

Push the ENTER button on the face of the alarm until the Set-Up Menu is

displayed.

2.

Scroll to SET UP and hit ENTER.

3.

The PASSWORD MENU displays (The Password is 0000). Press the

ENTER button repeatedly until the CAL CONFIG MENU is displayed.

4.

In the CAL CONFIG MENU set for the following: