Labconco Fiberglass 30 Laboratory Fume Hoods 3030005 User Manual

Page 19

Advertising
background image

15

Product Service 1-800-522-7658

Chapter 3: Getting Started


the model. All of the electrical connections are
terminated at the single point internal junction box for
hook-up by a qualified electrician. Refer to the wiring
diagram for your Fiberglass 30 Hood in Appendix C:
Fiberglass 30 Fume Hood Specifications.

Figure 3-2


All wiring for the fume hood
SHOULD be performed by a licensed
electrician and conform to all local
codes. In most cases, the hood will
require the use of shielded conduit to
protect the wiring into the hood. The
grounding connection shall not be
made to the terminal box cover.

The incandescent light has been mounted inside the top
liner panel and is sealed from vapors inside the hood
structure. To change the incandescent light bulbs in
your hood, you must first remove the protective glass.
Replace the defective bulbs, and reassemble.

Internal Junction
Box