Refer to the following table if your fume hood fails to
operate properly. If the suggested corrective actions do
not solve your problem, contact Labconco for
additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Remote blower and
light won’t operate

Wires not connected
at junction box or
switches.

Check connection of switches.

Check connection to control box on top
of unit.

Circuit

breakers

tripped in building
electrical supply.

Reset circuit breakers.

Remote blower
won’t operate, but
light works

Blower wiring is
disconnected.
Belt broken.

Inspect blower wiring and switch.

Replace belt.

Blower

motor

is

defective.

Replace blower motor.

Fume hood blower
operates but light
will not operate

Lamp not installed
correctly.

Inspect lamp installation.

Lamp is defective.

Replace lamp.

Circuit breaker in
building is tripped.

Reset the circuit breaker.