Chapter 3: Getting Started

Connecting the Service
Lines to the Fiberglass 30
Fume Hood

The hoods with service fixture kits are plumbed from
the valve to the hose connector. Supply tubing shall be
provided by the qualified installer. Tubing can enter
the hood from above, through the back, or through the
work surface to make these connections to the service
fixtures.


NOTE: Inspect all fittings for
leakage. Tighten the fittings slightly if
needed.






Should access to the hood plumbing fixture bodies be
required, remove the service access plate on the side of
the hood by loosening screws. The valve body will be
fully exposed for any service work that may be
necessary. The service fixture kits on your laboratory
hood are designed for use with the following services:

• Air

• Hot Water

• Vacuum

• Cold Water

• Natural Gas – See caution below




!

CAUTION: Do not use oxygen with any standard
service fixture. Contact Labconco Customer
Service for oxygen fixture information.

!

WARNING: Contact Labconco Customer
Service directly before using any service other
than those listed above in these valves to assure
full compatibility.

!