Xtramus NuDC-4U V2.1 User Manual

Page 29

Advertising
background image


29

The display scales for Voltage (V), Ampere (A), and (Watt) can be set with the scroll-down menu in these
fields. Display scales available for Voltage (V), Ampere (A), and (Watt) are shown in the table down
below:

Voltage (V)

100mV, 200mV, 500mV, 1V, 2V, 5V, 10V, 20V

Ampere (A)

100mA, 200mA, 500mA, 1A, 2A, 5A

Watt (W)

100mW, 200mW, 500mW, 1W, 2W, 5W, 10W, 20W, 50W, 70W, 90W

If you set the display scale of Voltage (V), Ampere (A), or (Watt) to a larger scale, the corresponding curve
displayed on the Main Display Screen will be smoother than the curves with smaller scale, as shown in
the figures down below:

Ch_A_V Curve: Set as 1V

Ch_A_V Curve: Set as 5V

You can choose specific curve displayed on the Main Display Screen. The curve you chose will be
thicker and highlighted in white as shown in the figures down below:

Ch_A_V Curve: Not chosen

Ch_A_V Curve: Chosen and Highlighted

To select a curve, please click the empty space in front of the curve you
would like to choose. An arrow icon

will appear when a curve is chosen.

You can move the curve you’ve chosen with Curve Moving Buttons as
well.

Also, you can choose which curve shall be displayed on the Main Display

Screen

by clicking the colored square in front of it, as shown in the figure at

right. The curve you’ve chosen this way will only be hidden. Tests regarding
to that curve will still be running. The colored square that represents the

curve will be replaced by a grave square icon

if it is hidden.

E-mail: [email protected]

Website: www.Xtramus.com

XTRAMUS TECHNOLOGIES

®

Advertising