Power-on conditions – AMETEK ReFlex Programming Manual User Manual

Page 116

Advertising
background image

ReFlex Power™ Programming Manual

Load Module Remote Programming

112

M380056-03 Rev L

R

EADBACK

F

UNCTIONS

•

Actual measured voltage, power and current

•

Voltage, current and resistance settings

•

Soft voltage, power and current limits

•

Status and Accumulated Status registers

•

Programming error codes

•

Fault codes

•

Manufacturer, power supply model, serial number, and firmware

version identification

5.3

POWER-ON CONDITIONS

Table 5-1 presents the power-on conditions for the Active Loads.

Table 5-1. Load Power-on Conditions

Condition

Default

Current

0 Amps (initial power–on current). Also see

CAL<n>:INIT:CURR

Soft Voltage Limit

Model maximum voltage

Soft Current Limit

Model maximum current

OVP Trip Voltage

Model maximum voltage +7% (initial power–on OVP).
See

CAL<n>:INIT:VOLT:PROT

Input

OFF, relays open

Grouped

No Modules Grouped (example paralleled)

Soft Power Limit

Model power limit