Table 2-2. SCPI Error Codes

Error Code

Description

0

No Error
The error queue is empty.

2

Invalid Index
When a command is issued to an empty slot

11

Wrong Module Type
Command is not correct for the specific module type.

12

Feature not supported
Specific command may not get implemented for the module type

14

Feature Not Implemented
Command will be implemented in the future

15

System too Busy
System is busy executing a previous command, the command causing this error will
not be executed.

206

TrigChannel not available
Requested trigger channel is already used in the system

251

Wrong Group Config/Oper
Group creation failed for stated reason.

252

Module inhibited asserted
Module Inhibit is asserted in the RFPC for the ReFlex Power™ system.

253

List Seq Buffer Full
The LIST being created, is too long.

-102

Syntax error
An unrecognized command or data type was encountered.

-200

Execution error
A command could not be executed due to the current condition of the module.

-222

Data out of range
Parameter exceeded range of valid values.

-292

"Name not found/invalid
Requested File name either not found or the name was not valid.

-293

Name already exists
Requested name already exists.

-331

System Error
A valid command could not be executed due to some internal/earlier error.

-350

Queue overflow
The error queue can contain up to ten entries. If more than ten error/event conditions
are logged before a SYSTem:ERRor? Query, an overflow will occur; the last queue
entry will be overwritten with error -350. When the queue overflows, the least recent
error/events remain in the queue and the most recent error/events are discarded.

2.8

LXI™ AND SCPI CONFORMANCE INFORMATION

The ReFlex Power™ Controller Ethernet is IEEE 802.3 and LXI™ (LAN eXtensions
for Instrumentation) class C compliant.

The syntax of all SCPI commands implemented by the ReFlex Power™ system
and documented in this manual, are either SCPI commands confirmed in the SCPI
1999.0 Specification, Volume 2: Command Reference, or they are customized
commands, not part of the SCPI definition, but do conform to SCPI syntax.