10. Problems and solutions

The machine does not switch on when the ON/OFF button is pressed.

What should I do?

(1) Make sure that the batteries have been inserted correctly and have contact.

(2) Change the batteries if necessary.

(3) Contact customer service.
The electrodes are detaching themselves from the body. What should I

do?

(1) Clean the adhesive surface of the electrodes with a damp, lint-free cloth.

Then let them dry in the air and attach them again. Should the electrodes

still not adhere firmly, they must be replaced.

(2) Before each application, clean the skin, avoiding using any skin cream or

oils before treatment. Shaving can increase adhesion of the electrodes.

The machine is giving out abnormal signal tones during treatment. What

should I do?

(1) Look at the display. Is any channel flashing? Interrupt the programme

by pressing the ON/OFF button. Make sure that the connecting leads

are properly connected to the electrodes. Make sure that there is a firm

contact between the electrodes and the treatment area.

(2) Make sure that the plug of the connecting lead is firmly connected to the

machine.

(3) If the signal tones do not stop and the channel is flashing, change the

connecting cable.

(4) The display shows a flashing battery signal. Renew the batteries.
There is no noticeable stimulation. What should I do?

(1) If a warning signal sounds, go through the steps described above.

(2) Press the ON/OFF key to start the programme again.

(3) Check that the electrodes are positioned properly and make sure that the

adhesive electrodes are not overlapping.

(4) Increase pulse intensity in stages.

(5) The batteries are almost empty. Replace these.

You are aware that the electrodes feel uncomfortable. What should I do?

(1) The electrodes are badly positioned. Check positioning and if necessary

reposition the electrodes.

(2) The electrodes are worn. Due to the fact that current distribution can no

longer be guaranteed evenly over the entire surface, these may lead to

skin irritation. You must therefore replace these.

The skin in the treatment area becomes red. What should I do?

Stop treatment immediately and wait until the skin has returned to normal. If

any redness beneath the electrode soon disappears, this is not dangerous

and is due to the increased circulation which has been stimulated locally.

If, however, the skin irritation remains and there is possibly itching or inflam-

mation, please consult your doctor before further use. It may possibly be

caused by an allergy to the adhesive surface.
The belt clip is not secure. What should I do?

If the clip is fastened to the belt and only then connected to the machine,

there is not optimal hold. You should therefore take the machine in your

hand. Position the belt clip on the back of the machine. You will notice the

correct position of the clip when you hear it click into position. Now secure

the machine with belt clip on the belt again.

11. Technical details

Name and model:

EM 41

Initial curve shape:

biphase square-wave pulse

Pulse duration:

40– 250 µs

Pulse frequency:

1–120 Hz

Output voltage:

max. 90 Vpp (at 500 Ohm)

Output current:

max. 180 mApp (at 500 Ohm)

Power supply:

3x AAA batteries

Treatment time:

adjustable from 5 to 90 minutes

Intensity:

adjustable from 0 to 15

Operating conditions: 10 °C – 40 °C (50 °F – 104 °F) with a relative humidity

of 30 – 85%

Storage conditions:

-10 °C – 50 °C (14 °F – 122 °F) with a relative humidity

of 10 – 95%

Dimensions:

122 x 59 x 23 mm