Flowserve ValveSight Diagnostics User Manual

Page 55

Advertising
background image

ValveSight Diagnostics for HART User Manual FCD VSENSF0001-02-AQ 03/15

55

flowserve.com

#

Indicator

Implications

Possible

Solutions

Default

Settings

Enable/

Disable

Refer to

DTM

Screen

Blink

Code

36

Watch Dog

Time Out

Alarm

Indicates that functions in the

firmware took longer than expected.

This can be caused when intermittent

operation occurs when connecting

power.

Remove power and then reconnect to

clear this alarm. If the alarm persists

replace the main circuit board.

N/A

Yes

Health >

Positioner

Health

RRRY

37

Non-Volatile

Ram Check-

sum Alarm

Indicates the check sum of the internal

data was not updated correctly. This

may affect the function of the posi-

tioner in various ways or not at all.

Error may clear with time. If error

persists, cycle power and complete a

QUICK-CAL. If the error still persists,

replace the main circuit board.

N/A

Yes

Health >

Positioner

Health

RRRY

38

N/A

39

Electronic In-

ability to Fail

Safe Warning

Indicates that the peizo may be dam-

aged. This may prevent the proper

failure position upon loss of signal/

power. This condition may occur

briefly on an air-to-close valve that is

held for long periods of time in the

closed position, or and air-to-open

valve held in the open position.

If alarm persists for more than 30

minutes, the piezo assembly is

damaged and should be replaced.

N/A

Yes

Health >

Positioner

Health

YRRY

40

Supply Pres-

sure Low

Alarm

Indicates that the supply pressure is

below the user set alarm limit. Low

supply pressure can cause poor valve

response or positioner failure. The

minimum recommended supply pres-

sure is 30 PSI (2.1 bar) for proper

operation. The unit will fail at less

than approximately 17 PSI (1.2 bars).

Low supply pressure indicates can

also be caused by pneumatic leak.

Regulate the supply pressure at the

positioner above 30 PSI (2.1 bar).

Recalibrate pressure sensors. Ensure

system air/gas supply is adequate.

Repair kinked supply tubing. Check the

pressure sensor board connections

and replace pressure sensor board if

necessary. Check for pneumatic leaks

in the actuator and actuator tubing.

22 PSI

Yes

Health >

Actuator

Health

RYYG

41

Supply Pres-

sure High

Warning

Indicates the positioner has deter-

mined that the supply pressure is

above the user set warning limit.

Regulate the supply pressure at

the positioner below the maximum

limit recommended for your actuator.

Recalibrate pressure sensors. Check

the pressure sensor board connec-

tions. Replace pressure sensor board

if necessary.

150 PSI

Yes

Health >

Actuator

Health

YYGR

42

Supply Pres-

sure Low

Warning

Indicates the supply pressure is be-

low the user set warning limit. Lows

supply pressure can cause poor valve

response or positioner failure. The

minimum recommended supply pres-

sure is 30 PSI (2.1 bar) for proper

operation. The unit will fail at less

than approximately 17 PSI (1.2 bars).

Low supply pressure indicates can

also be caused by pneumatic leak.

Regulate the supply pressure at the

positioner above 30 PSI (2.1 bar).

Recalibrate pressure sensors. Ensure

system air/gas supply is adequate.

Repair kinked supply tubing.

Check the pressure sensor board

connections and replace pressure

sensor board if necessary. Check for

pneumatic leaks in the actuator and

actuator tubing.

30 PSI

Yes

Health >

Actuator

Health

YYYG

43

Pressure

Port 1 Out

of Range

Warning

Indicates the internal pressure sensor

at Port 1 is either saturated with a

pressure over 150 PSI (10.3 bar) or

the sensor connection has failed.

Check the supply pressure. Calibrate

pressure sensors. Check the pressure

sensor board connections and replace

pressure sensor board if necessary.

N/A

Yes

Calibration

> Actuator/

Stroke/

Friction

YYGY

44

Pressure

Port 2 Out

of Range

Warning

Indicates the internal pressure sensor

at Port 2 is either saturated with a

pressure over 150 PSI (10.3 bar) or

the sensor connection has failed.

Check the supply pressure. Calibrate

pressure sensors. Check the pressure

sensor board connections and replace

pressure sensor board if necessary.

N/A

Yes

Calibration

> Actuator/

Stroke/Fric-

tion

YYGY

45

Pressure

Port 1 Range

Small Warn-

ing

Indicates that during a pressure sen-

sor calibration, the range of applied

pressures to Port 1 was too small for

optimum performance.

Adjust the supply pressure to a higher

value so the positioner can properly

span the sensors, then recalibrate.

Briefly push the QUICK-CAL button

to acknowledge this condition and

the positioner will operate using the

current short stroke calibration values,

if valid.

N/A

No

Calibration

> Actuator/

Stroke/Fric-

tion

YYGY

Advertising