General Information

Getting Started

1

1-11

5. Press the UNITS SELECT button if an integrated exposure is to be measured. Or, if exposure rate

is desired, these same units may be chosen with a timebase of seconds, minutes or hours.
Additionally, per frame and amperes are available. Select desired units of measurement by
pressing the UNITS SELECT button. Change the selection by pressing the UP or DOWN buttons.
If exposure/frame units are selected, the user must set the frame rate at this screen.

6. If an ion chamber using air density corrections is selected, press the AIR DENSITY button to

display the Air Density screen. If the temperature & pressure mode has been set to Automatic
Sense, the measured ambient temperature and pressure will be displayed. If the temperature &
pressure mode has been set to Manual Entry, the last used temperature and pressure will be
displayed. Set temperature and

pressure to match ambient conditions.


When air density corrections are in effect, a “ADC”
annunciator will appear on the Measurement
screen. In Automatic Sense mode the ambient
temperature and pressure are updated each time
RESET/MEASURE is pressed.

7. Press and release the RESET/MEASURE button to place the dosimeter in the measurement mode.

8. Make the exposure. Following an exposure, the display will be updated with measurement results

for the exposure. Additional exposures may be made at this point as the dosimeter will detect each
exposure and automatically update the display.


Complete detailed operation concerning the Model 35050AT Dosimeter front panel operation may be
found in section 2 of this manual, including the use of the Model 35080/80A/80B kVp Divider. If you wish
to control these functions over the RS-232 Interface Cable, consult the Model 35050A Programmer's
Instruction Manual.