4 fuse replacement – Fluke Biomedical 10100AT User Manual

Page 56

Advertising
background image

10100AT
Operators Manual

4-2



















Figure 4-1. The Model 35050AT Dosimeter Battery Replacement

4.4 Fuse Replacement

1. Twist off the four rubber feet on the bottom of the instrument with finger friction. Remove the

bottom cover plate and set aside.

2. Invert the instrument and tap into the palm of the hand till the battery pack falls out.

3. Look inside the battery pack compartment and find the fuse extractor. It is a loop of plastic tubing

held captive by the fuse body & fuse clips (see Figure 4-2). Grasping the fuse extractor tightly
between your index finger and thumb and as close to the fuse as possible, push away from you
until the fuse pops out. Pull the fuse out of the instrument.


















Figure 4-2.

The Model 35050AT Dosimeter Fuse Replacement

4. Replace the fuse with P/N FU-83.

RUBBER FEET

RUBBER FEET

DOSIMETER (REAR VIEW)

BOTTOM COVER PLATE

FUSE

FUSE

EXTRACTOR