A.3 Model 96035B (15 cc) Ion Chamber

The standard calibration for the 15 cc Model 96035B ion chamber includes one mammographic point (the
Fluke Biomedical, Radiation Management Services equivalent of NIST beam quality MoMo28) and one
diagnostic point (the Fluke Biomedical, Radiation Management Services equivalent of NIST beam quality
L100). Each of these calibration factors in entered into the Model 35050AT Dosimeter and linked to two
Ion Chamber Selection screens (for a total of four screens) that read as follows:

15 cc xxxxx MAM ADC

15 cc xxxxx MAM

15 cc xxxxx DIA ADC

15 cc xxxxx DIA

15 cc refers to the volume of the Model 96035B ion chamber.

xxxxx is replaced in your selection by the serial number of your Model 96035B.

MAM indicates the selection for the mammographic calibration point (MoMo28).

DIA indicates the selection for the diagnostic calibration point (L100).

ADC indicates that air density correction are active.

The selection containing MAM should be used for all measurements on the mammographic side of the
chamber, while the selection containing DIA should be used for all measurements on the diagnostic side
of the chamber. The selections containing ADC indicate that automatic air density corrections will be
applied, while the selections without ADC indicate that no air density corrections will be applied.