2.8 Exclusive Measurement Mode

The firmware allows a Model 35050AT Dosimeter to be setup to perform exclusive exposure and kVp
measurements. This exclusive option can be enabled through the Customization Software. When a
Model 35050AT Dosimeter configured to perform exclusive measurements, the operator selects whether
an exposure or kVp measurement is to be made from the dosimeter's front panel.

The Model 35050AT Dosimeter contains the filter pack type “35080/80A Not Used”. This filter pack
selection turns off the Model 35050AT Dosimeter's kVp divider port. The Model 35050AT Dosimeter
ignores the 35080 Input when this filter pack is selected, and displays only the exposure line on the
measurement screen.













Selecting any one of the other filter packs causes the dosimeter to monitor the kVp divider port and
present only the kVp line on the measurement screen: