10100AT
Operators Manual

4-4

Use of a different fuse could cause damage to the
instrument.

5. Install the new fuse with the fuse extractor into the instrument by reversing the procedure.

4.5 Replaceable Parts Information

Fluke Biomedical, Radiation Management Services maintains a complete inventory of all normal
replaceable parts (see Table 4-1). To place an order, or to obtain information concerning replaceable
parts, contact the Fluke Biomedical at 440.248.9300.

Table 4-1. Replaceable Parts List

Description Part

Number

20 ft (6.1 m) Coax/Triax Cable

38208

Ion Chamber Stem

32147

Battery Holder

87-3

HVL Filter Kit

37668

Carrying Case, Fitted

37500D

Chamber Holder

37582

HVL Filter Tray

37586

Test Stand Rod

37589

Test Stand Base

37590

Thumbscrew 2715-E-BLACK

Rubber Feet

38471

Bottom Cover Plate

34067

Fuse

19-5

Fuse Extractor

38486

Dust Cap and Chain

878-11-63

Customization Kit with 2 m Cable & Adapters

37594

A.C. Adapter

Call for P/N

4.6 Temperature / Pressure Sensor Adjustment

The measurements from the internal temperature / pressure sensor may be adjusted using the front panel
buttons. Allow the unit’s temperature to equilibrate for at least 1 hour before calibrating the sensor. Verify
that only calibrated thermometers and barometers are used to adjust the 35050AT internal temperature /
pressure sensor. To adjust the sensor measurements:

1. Make sure that the Auto Sense Temperature & Pressure option has been selected.

2. Momentarily depress the Options and Air Density buttons simultaneously.

3. The adjusted/measured temperature and adjusted/measured pressure will be displayed and the

cursor (>) appears next to the value to be adjusted.

CAUTION