4. Momentarily depressing the Options and Air Density buttons simultaneously again toggles the

cursor to the other value. Holding down the Options and Air Density buttons simultaneously
causes the cursor to toggle between temperature and pressure values.

5. Press the Up or Down buttons to increment or decrement the adjusted temperature or pressure

readings until they agree with the calibrated thermometer or barometer. Setting the adjusted
values equal to the measured values restores the adjusted values to the measured sensor
values.

6. Pressing any other button exits the adjustment mode.

4.7 Recalibration and Repair Information/Technical

Assistance

If your instrument needs recalibration or repair, we request that you consult the Fluke Biomedical,
Radiation Management Services Repair Department at 800.850.4606, 440.498.2564, or Fax
440.542.3682.

More information concerning the operation, application, or service of your instrument may be obtained
from the applications engineer at the numbers listed above.

4.8 Warranty Information

Should your instrument require warranty service, contact Fluke Biomedical at 440.248.9300.



This warranty does not cover battery replacement
or damage caused by battery leakage.