Gas Flow Analyzer

Controls and Indicators

7

Table 4. Controls and Indicators

No. Name

Description/Use

A

Miniature RS-232 Serial Port

Connect the RS-232 cable here (for use with VT for Windows
software.)

B

Battery Eliminator Connector

Connect the battery eliminator here. Use the battery eliminator
whenever possible.

C

Battery Eliminator Indicator

Glows green when the battery eliminator is supplying power to the
Analyzer. Note that this power source cannot recharge the battery.

D

High Pressure Connector

Connect the High Pressure Adapter here. The Analyzer makes high
pressure measurements from -2 to 100 psi.

E

Flow and Low Pressure Connector

Insert one of three coded modular connectors here; the Analyzer
automatically detects the connector type. The Analyzer use the
High-Flow Sensor to measure up to

± 150 lpm, the Low-Flow

Sensor to measure up to

± 25 lpm, or the Low-Pressure Adapter to

measure from -20 to 120 cmH2O.

F

Display Screen

Shows measurement parameters and statistics (Min, Avg, Max),
waveforms, and setup selections. English only.

G

Oxygen Sensor Connector

Connect the Oxygen Sensor here. The Analyzer makes O2
measurements from 0 to 100%.

H

Temperature, RH Sensor Connector

Connect the optional Temperature and Relative Humidity Sensor
here.