Gas Flow Analyzer

Specifications

29


Maximum Flow Rate
(absolute value):


35 lpm

Operating Flow
Range:


± 25 lpm

Accuracy:

± 3 % of reading or ± 1% of
range, whichever is greater

Floor for Absolute
Accuracy:


3 lpm

Resolution:

0.01 lpm flow > 1 lpm

Freq. Response:

> 25 Hz or t

10-90

< 40 ms

Sample Rate:

100 Hz

Dynamic
Resistance:


< 2.5 cmH

2

O @ 5 lpm

Low-Flow Dropout:

0.24 lpm

Breath-Detect
Threshold:


1 lpm (user selectable)

Volume Range:

± 60 l

Volume Accuracy:

± 3 % of reading or ± 5 ml


Oxygen Measurement

Range:

0 to 100 %

Accuracy:

± 2 % full-scale output

Resolution:

0.1 % O

2

Frequency
Response:


> 15 s (t

10-90

)

Sample Rate:

100 Hz

Sensor
Technology:


Galvanic Fuel Cell

Calibration:

Allows user calibration using air
and 100 % O

2

Notes:
• Automatic partial pressure compensation for
barometric and airway pressure changes
• Recommended interval for changing oxygen sensor is
one year. However, sensor may last longer. During user
calibration of the sensor, the Analyzer can detect if the
sensor needs to be replaced.