Gas Flow Analyzer

Measuring Low Flow

17

Measuring Low Flow

Measure low flow in the range: +/- 25 lpm. Never exceed
an absolute maximum of 35 lpm.

Refer to Figure 5 while following these steps:

1.

Press R and follow on-screen instructions.

2.

At the Analyzer, connect the Low-Flow Sensor to the

Flow/Low-Pressure Port.

3.

Press K (in OTHER MEASUREMENTS) or N

(in WAVEFORM SCREENS.)

4.

Connect the Low-Flow Sensor to the source.

5.

Apply flow.

6.

Successively press G (Units) to select a flow unit

in the sequence: LPM

Æ CFM Æ LPS Æ ml/min Æ

ml/sec.

If the waveform is too small or too large, press F
(Rescale) to view the waveform within optimized limits.

F1

F2

Back

EN

TE

R

Setup

Mem

ory

Mem

ory

VE

NT

ILA

TO

R P

AR

AM

ET

ER

S

FL

OW

/VO

L

FL

OW

/VO

L

TREND

TEST

ZER

O

SENSORS

MORE

PRESSURE

PRESSURE

PRESSURE

TIM

IN

G

VO

LU

M

E

FL

OW

W

AV

EF

OR

M S

CR

EE

NS

OT

HE

R M

EA

SU

RE

ME

NT

S

SP

EC

IA

L FU

NC

TIO

NS

%RH

O2

_ , @

_ , @

11

22

ABC

ABC

55

JKL

JKL

44

GHI

GHI

66

MNO

MNO

88

00

TUV

TUV

77

PQRS

PQRS

99

WXYZ

WXYZ

33

DEF

DEF

GA

S F

LO

W

A

NA

LY

ZE

R

GA

S

FL

OW

A

NA

LY

ZE

R

VT MOBILE

VT MOBILE

Low Flow

Sensor

ede13f.eps

Figure 5. Measuring Low Flow

Advertising