Checking ventilator parameters – Fluke Biomedical VT Mobile Getting Started User Manual

Page 30

Advertising
background image

VT MOBILE
Getting Started Manual

20

For waveform pressure readings, press F (Rescale) to
optimize waveform viewing.

WWarning

Always turn pressure off at the source prior
to disconnecting either end of the High-
Pressure Adapter. Serious injury could
otherwise result.

Figure 0-8 shows some characteristic high pressure
screens.

OTHER

MEASUREMENTS

WAVEFORM

SCREENS

ede04f.eps

Figure 8. High Pressure Screens

Checking Ventilator Parameters

Refer to Figure 9 when you are connecting the Analyzer
to a ventilator and test lung.

1.

Press R and follow on-screen instructions.

2.

At the Analyzer, connect the High-Flow Sensor to the

Flow/Low-Pressure Port.

3.

Connect the Oxygen Sensor to the upper port on the

right side of the Analyzer.

4.

If available, connect the optional Temperature/RH

Sensor to the lower port on the right side of the
Analyzer.

5.

Using parts from the Accessory Kit and the

connectors provided with the sensors, make
connections between the ventilator and the
ACCU LUNG as shown in Figure 9. Use the
sequence: ventilator Y connector

Æ High-Flow

Sensor (blue stripe toward the ACCU LUNG)

Æ

Temperature/RH Sensor

Æ Oxygen Sensor Æ

ACCU LUNG. Vertically align all sensors.

6.

Set up the ventilator for a characteristic breath

pattern. For example, you could set 10 BPM at 7.5
lpm.

7.

Set the ACCU LUNG for C20 compliance (both outer

springs engaged) and Rp50 resistance (positioned as
shown in Figure 9.)