Inspiratory, expiratory time, Peak inspiratory pressure (pip), Inspiratory pause pressure (ipp) – Fluke Biomedical VT Mobile Getting Started User Manual

Page 41: Mean airway pressure (map), Positive-end expiratory pressure (peep), Peak expiratory flow (pef)

Advertising
background image

Gas Flow Analyzer

Specifications

31

Inspiratory, Expiratory Time

Resolution:

0.01 s

Range:

0.25 to 9.99 s

Accuracy:

± 2 % or 0.1 s

Peak Inspiratory Pressure (PIP)

Resolution:

0.1 cmH

2

O

Range:

± 120 cmH

2

O

Accuracy:

± 3 % or 2 cmH

2

O

Inspiratory Pause Pressure (IPP)

Resolution:

0.1 cmH

2

O

Range:

± 120 cmH

2

O

Accuracy:

± 3 % or 2 cmH

2

O

Mean Airway Pressure (MAP)

Resolution:

0.1 cmH

2

O

Range:

± 80 cmH

2

O

Accuracy:

± 3 % or 1 cmH

2

O

Positive-End Expiratory Pressure (PEEP)

Resolution:

0.1 cmH

2

O

Range:

- 5 to 40 cmH

2

O

Accuracy:

± 3 % or 1 cmH

2

O

Peak Expiratory Flow (PEF)

Resolution:

0.01 lpm

Range:

0 to 150 lpm

Accuracy:

± 3 % or 1 lpm

Advertising