VT MOBILE
Getting Started Manual

22

8.

Apply power to all elements of the test setup. For the

Analyzer, press

H

on, wait for the Zeroing screen,

and then press

G

. The Analyzer is ready to use

once the Tidal Volume screen appears.

9.

Press the access keys shown in Table 5 to display

the various ventilator parameters. Once you have
pressed the first key, you can continue pressing that
key to view other parameters or press

G

to view

all the parameters.

As described in Table 5, the Analyzer calculates 16
breath parameters in Local Mode. Note that all 16
parameters and other information can be viewed on one
screen in Remote mode when using the VT for Windows
software.

Figure 10 shows some characteristic ventilator parameter
screens.

Figure 11 shows additional oxygen, temperature, relative
humidity, and barometric pressure parameter screens.

Figure 12 shows additional airway pressure measurement
screens accessible during the ventilator check.

Table 5. Ventilator Parameters

Keys Item

Name

VT Tidal

Volume

MV Minute

Volume

PIF

Peak Inspiratory Flow

K

G

PEF

Peak Expiratory Flow

PIP

Peak Inspiratory Pressure

PEEP Positive End-Expiratory Pressure

MAP

Mean Airway Pressure

L

G

IPP

Inspiratory Pause Pressure

Rate Breath

Rate

Ti Inspiratory

Time

Te Expiratory

Time

M

G

I:E

Inspiratory to Expiratory Time Ratio

P

O2 Oxygen

Percentage

Æ P

Temp Temperature (gas under test)

Æ P

RH

Relative Humidity (gas under test)

S E

Baro Barometric

Pressure