Gas Flow Analyzer

Specifications

27

Measurement Specifications

Low-Pressure Port

Maximum Applied
Pressure:


5 psi

Operating Pressure
(Differential):


-20 to 120 cmH

2

O

Operating Pressure
(Common Mode):


NA

Span Accuracy:

± 2 % of reading or 1.5
mmHg, whichever is
greater

Frequency Response:

> 10 Hz

Resolution:

0.1 mmHg

Sample Rate:

100 Hz

Fittings:

Flow Connector with 2
tubes "T" connected to a
single fitting

Note: No fluid may be applied to port.

High-Pressure Port

Maximum Applied
Pressure:


125 psi

Operating Pressure:

- 2 psi to 100 psi

Span Accuracy:

± 2 % of reading or ± 0.2
psig, whichever is greater

Frequency
Response:


> 10 Hz

Resolution:

0.1 psi

Sample Rate:

100 Hz

Fittings:

Single port, Luer lock,
stainless steel

Note: No fluid may be applied to port.