Measurement specifications, Low-pressure port, High-pressure port – Fluke Biomedical VT Mobile Getting Started User Manual
Page 37

Gas Flow Analyzer
Specifications
27
Measurement Specifications
Low-Pressure Port
Maximum Applied
Pressure:
5 psi
Operating Pressure
(Differential):
-20 to 120 cmH
2
O
Operating Pressure
(Common Mode):
NA
Span Accuracy:
± 2 % of reading or 1.5
mmHg, whichever is
greater
Frequency Response:
> 10 Hz
Resolution:
0.1 mmHg
Sample Rate:
100 Hz
Fittings:
Flow Connector with 2
tubes "T" connected to a
single fitting
Note: No fluid may be applied to port.
High-Pressure Port
Maximum Applied
Pressure:
125 psi
Operating Pressure:
- 2 psi to 100 psi
Span Accuracy:
± 2 % of reading or ± 0.2
psig, whichever is greater
Frequency
Response:
> 10 Hz
Resolution:
0.1 psi
Sample Rate:
100 Hz
Fittings:
Single port, Luer lock,
stainless steel
Note: No fluid may be applied to port.