Airway pressure, High-flow port – Fluke Biomedical VT Mobile Getting Started User Manual
Page 38
VT MOBILE
Getting Started Manual
28
Airway Pressure
Maximum Applied
Pressure:
5 psi
Operating Pressure:
- 20 cmH
2
O to 120 cmH
2
O
Span Accuracy:
± 2 % of reading or ± 2.0
cmH
2
O, whichever is greater
Frequency
Response:
> 25 Hz or t
10-90
< 40 ms
Resolution:
0.1 cmH
2
O
Sample Rate:
100 Hz
Fittings:
Internally connected to flow
sensor pressure lines
High-Flow Port
Maximum Flow Rate
(absolute value):
200 lpm
Operating Flow Range:
± 150 lpm
Accuracy:
± 3 % of reading or ± 2 %
of range, whichever is
greater
Floor for Absolute
Accuracy:
25 lpm
Resolution:
0.01 lpm
Frequency Response:
> 25 Hz or t
10-90
< 40 ms
Sample Rate:
100 Hz
Dynamic Resistance:
< 2 cmH
2
O @ 60 lpm
Low-Flow Dropout:
2.5 lpm
Breath-Detect
Threshold:
4 lpm
Volume Range:
> ± 60 l
Tidal Volume Accuracy: ± 3 % or reading or ± 20
ml, whichever is greater