Airway pressure, High-flow port – Fluke Biomedical VT Mobile Getting Started User Manual

Page 38

Advertising
background image

VT MOBILE
Getting Started Manual

28

Airway Pressure

Maximum Applied
Pressure:


5 psi

Operating Pressure:

- 20 cmH

2

O to 120 cmH

2

O

Span Accuracy:

± 2 % of reading or ± 2.0
cmH

2

O, whichever is greater

Frequency
Response:

> 25 Hz or t

10-90

< 40 ms

Resolution:

0.1 cmH

2

O

Sample Rate:

100 Hz

Fittings:

Internally connected to flow
sensor pressure lines

High-Flow Port

Maximum Flow Rate
(absolute value):


200 lpm

Operating Flow Range:


± 150 lpm

Accuracy:

± 3 % of reading or ± 2 %
of range, whichever is
greater

Floor for Absolute
Accuracy:


25 lpm

Resolution:

0.01 lpm

Frequency Response:

> 25 Hz or t

10-90

< 40 ms

Sample Rate:

100 Hz

Dynamic Resistance:

< 2 cmH

2

O @ 60 lpm

Low-Flow Dropout:

2.5 lpm

Breath-Detect
Threshold:

4 lpm

Volume Range:

> ± 60 l

Tidal Volume Accuracy: ± 3 % or reading or ± 20

ml, whichever is greater