Barometric pressure measurement, Temperature and relative humidity, Measurement parameter specifications – Fluke Biomedical VT Mobile Getting Started User Manual
Page 40: Inspiratory and expiratory tidal volume (vt), Expiratory minute volume (mv), Breath rate (bpm)
Advertising

VT MOBILE
Getting Started Manual
30
Barometric Pressure Measurement
Range:
8 to 18 psia (400 to 900 mmHg)
Accuracy:
± 2 % of reading
Resolution:
0.1 mmHg
Frequency
Response:
< 5 s (t10-90)
Sample Rate:
N/A
Calibration:
Not required; however, device
allows user calibration of offset.
Temperature and Relative Humidity
Temperature Relative
Humidity
Resolution:
0.1
°
0.1 %
Range:
0 – 50
°C
10 to 95 %
Accuracy: ± 1 °C ± 10 % RH
Units:
°C, °F, °K
%
Measurement Parameter Specifications
Inspiratory and Expiratory Tidal Volume (VT)
Resolution:
0.1 ml
Range:
> 10 l
Accuracy:
± 3 %
Expiratory Minute Volume (MV)
Resolution:
0.001 lpm
Range:
0 to 60 l
Accuracy:
± 3 %
Breath Rate (BPM)
Resolution:
0.1 bpm
Range:
2 to 150 bpm
Accuracy:
± 1 %