Barometric Pressure Measurement

Range:

8 to 18 psia (400 to 900 mmHg)

Accuracy:

± 2 % of reading

Resolution:

0.1 mmHg

Frequency
Response:


< 5 s (t10-90)

Sample Rate:

N/A

Calibration:

Not required; however, device
allows user calibration of offset.

Temperature and Relative Humidity

Temperature Relative

Humidity

Resolution:

0.1

°

0.1 %

Range:

0 – 50

°C

10 to 95 %

Accuracy: ± 1 °C ± 10 % RH
Units:

°C, °F, °K

%

Measurement Parameter Specifications

Inspiratory and Expiratory Tidal Volume (VT)

Resolution:

0.1 ml

Range:

> 10 l

Accuracy:

± 3 %

Expiratory Minute Volume (MV)

Resolution:

0.001 lpm

Range:

0 to 60 l

Accuracy:

± 3 %

Breath Rate (BPM)

Resolution:

0.1 bpm

Range:

2 to 150 bpm

Accuracy:

± 1 %