Measuring high pressure – Fluke Biomedical VT Mobile Getting Started User Manual

Page 28

Advertising
background image

VT MOBILE
Getting Started Manual

18

Figure 6 shows some characteristic low flow screens.

OTHER

MEASUREMENTS

WAVEFORM

SCREENS

ede15f.eps

Figure 6. Low Flow Screens

Measuring High Pressure

The Analyzer can measure high pressure from -2 to 100
psi. Be certain of your possible source pressure before
measuring high pressure. For example, normal bulkhead
outlet pressure could be too high. Unregulated cylinder
pressure could also be too high.

Refer to Figure while following these steps:

1.

Press R and follow on-screen instructions.

2.

At the Analyzer, connect the High-Pressure Adapter

to the High-Pressure Port.

3.

Press L in the OTHER MEASUREMENTS or

WAVEFORM SCREENS section.

4.

Verify that the screen displays High Pressure (Phi) or

HPress. If the Low-Pressure Adapter is also
connected, you may need to press L a second
time.

5.

With the high pressure source set to zero output,

connect the High-Pressure Adapter to the source.
Note that various connection schemes are available.

6.

Apply pressure.

7.

Successively press G (Units) to select a pressure

unit in the sequence: PSI

Æ kPa Æ Bar Æ mBar Æ

ATM

Æ inH2O Æ InHg Æ cmH2O Æ mmHg.