GBS Elektronik MCA166-USB User Manual

Page 18

Advertising
background image

MCA166-USB

6.1.3 Power supply

Power supply adequate for commonly used NaI, HPGe, CdTe and proportional
counters

Current limited mode with indication of overload condition through software; short
circuit protected and automated recovery after overload or shortening

6.1.4 Preamplifier Voltages

± 12V; maximum current 60mA for each of the voltages

± 24V; maximum current 60mA for each of the voltages

Maximum total power at least 2W

6.1.5 High Voltage

Single positive or negative modular HV supply reaching from 50V to 3000V

Indication of high voltage sign and value by software

Calibration within 1% at maximum of bias voltage

Maximum current 0.5mA or 0.25W (e.g. 1000V and 0.25mA)

6.1.6 Battery

Rechargeable Li-ion battery with no memory effect

Deep discharge protected

Battery capacity 32Wh at 20°C

Charger input: 9V ... 14V / 0.8A

Measured values of the battery life time

If no detector connected: 30 hours

Expected battery life time with detection systems connected:

Standard IAEA NaI detector (ORTEC/SCIONIX): 27 hours

Bias:

500V, 100µA

Preamplifier:

100mW

Miniature CdTe detection probe (SDP310, RITEC, Latvia): 25 hours

Bias:

400V, 100nA

Preamplifier:

+12V/20mA, -12V/9.5mA (@8.4V)

Standard planar HPGe detector (e.g. Canberra, Ortec): 14 hours

Bias:

2000V, 1nA

Preamplifier:

±12V: 35/-25mA; ±24V: 55/-27mA (@8.4V)

18