MCA Hardware Setup Commands:

Command name

CMD_SET_GAIN

Description

The command sets the amplifier coarse(cg) and fine gain(fg).
cg:

2, 5, 10, 20, 50, 100, 200, 500 or 1000

fg:

5000 ... 15000 (for cg = 1000 fg: 5000 ... 10000)

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

cg

fg

end flag

Byte String (HEX)

A5

5A

4C

00

cg

l

cg

h

fg

l

fg

h

00

00

B9

9B

Used setup value

None

Affected setup value

coarse_gain, fine_gain

MCA state

STATE_READY

Command name

CMD_SET_INPUT_POLARITY

Description

The command sets the input polarity for the internal amplifier.
ip = 0 positive input signals

ip = 1 negative input signals

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

ip

0

end flag

Byte String (HEX)

A5

5A

56

00

ip

00

00

00

00

00

B9

9B

Used setup value

None

Affected setup value

mca_input_pol

MCA state

STATE_READY

Command name

CMD_SET_MCA_INPUT

Description

The command sets the ADC input to amplifier or to direct.
ip = 0 input amplifier

ip = 3 input direct (0 ... +3V, PUR OFF)
ip = 4 input direct (0 ... -3V, PUR OFF)

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

ip

0

end flag

Byte String (HEX)

A5

5A

54

00

ip

00

00

00

00

00

B9

9B

Used setup value

None

Affected setup value

mca_input_adc

MCA state

STATE_READY