MCA Measurement Setup Commands:

Command name

CMD_SET_ADC_RES_DISCR

Description

The command set the ADC resolution(res: 128, 256 … 4096), the LLD(Low level
discriminator) and the ULD(Upper level discriminator) to the values of its three

parameter.
The command will be ignored and responded with a error messages, if one ore

more parameter are invalid or a measurement is still running.

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

res

LLD, ULD

end flag

Byte String (HEX)

A5

5A

46

00

res

l

res

h

LLD

l

LLD

h

ULD

l

ULD

h

B9

9B

Used setup value

None

Affected setup value

channels, lld, uld, stab_state

MCA state

STATE_READY

Command name

CMD_SET_PRESETS

Description

The command sets the choice(pre)
PRESET_NONE = 0

PRESET_REAL = 1
PRESET_LIVE

= 2

PRESET_INT

= 3

PRESET_AREA = 4

and value(val) of the preset. The parameter val must be lower then 65536 for
PRESET_REAL and PRESET_LIVE! If a measurement is running, only val can be

changed.

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

pre

val

end flag

Byte String (HEX)

A5

5A

48

00

pre

l

pre

h

val

l

...

...

val

h

B9

9B

Used setup value

None

Affected setup value

presets, preset_value

MCA state

STATE_READY

Command name

CMD_SET_ROI

Description

The command sets the begin and end of the ROI for the preset integral and
area.(LLD <= begin < end and LLD < end <= ULD)

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

beg

end

end flag

Byte String (HEX)

A5

5A

49

00

beg

l

beg

h

end

l

end

h

00

00

B9

9B

Used setup value

LLD, ULD

Affected setup value

roi_begin, roi_end

MCA state

STATE_READY