Name

MMCA_SET_FAST

Description

The command sets the fast discriminator threshold.

1. Parameter

Threshold

integer value

0 ... 2499, default 400

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_SET_SLOW

Description

The command sets the slow discriminator threshold.

1. Parameter

Threshold

integer value

0 ... 2499, default 400

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_SET_PZC_MANUAL

18

Description

The command sets the PZC value and returns the PZC offset.

1. Parameter

PZC value

integer value

0 ... 2499

Result array

Byte offset 128

Number of measured pulses

integer value

Result array
Byte offset 130

averaged negative offset of
measured input pulses

integer value

Return Code

see MMCA_RESET command

Name

MMCA_SET_PZC_VALUE

19

Description

The command sets the PZC value.

1. Parameter

PZC value

integer value

0 ... 2499

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_SET_PZC_TIME_OFFSET

Description

The command sets the time offset for the PZC measurement.

1. Parameter

Offset for low DTC

integer value

0 ... 31, default 15

2. Parameter

Offset for high DTC integer value

0 ... 31, default 15

Results

-

Return Code

see MMCA_RESET command

18 Please note that this function may cause a firmware hang up if the following requirements are not

meets:
- statistical distributed input pulses with a count rate between 500 and 30000 Cps

- the significant peak is located above the first quarter of the spectrum

19 The function requires a reference to a pzc result array, but the function has no effect on this array at

the moment.