Name

MMCA_QUERY_VOLTAGE_CURRENT

Description

The command reads the voltages and currents from the MMCA.

Parameter

-

Results array

Charger current [mA]

long

Byte Offset 4

HV primary current [mA]

long

Byte Offset 8

Battery current [mA]

long

Byte Offset 12

Battery voltage [mV]

long

Byte Offset 16

HV reference voltage [V]

long

Byte Offset 20

HV control voltage [V]

long

Byte Offset 24

+12V primary current [mA]

long

Byte Offset 28

+24V primary current [mA]

long

Byte Offset 32

-24V primary current [mA]

long

Byte Offset 36

-12V primary current [mA]

long

Return Code

see MMCA_RESET command

A3.3 MS-Windows DLL for the communication with several MCA166-USB

This Dynamic Link Library allows the communication with several MCA166-USB. It can manage the serial

communication ports COM1 to COM20.

Die Dynamic Link Library exists as a 16-bit variant (mca_comx.dll) and as a 32-bit variant
(mca32cmx.dll). The DLL must reside in the system directory of Windows or in the same directory as the

calling application. In order to include the DLL in your program, you have to use the in Visual Basic the
file MCA32CMX.BAS (valid only for Win32 applications) or in C/C++ the files MCA_COMX.H and

MCA_COMX.LIB or MCA32CXM.LIB. All functions and structures are declared in these files.

The use of the function is analogous to the functions of the DLLs for the comunication with one MCA166-
USB (see chapter A3.1). However, following differences exist: The return value of the function COMM_INIT

is an index of the communication port. You have to pass this index to each function as first parameter so
that you can distinguish the communication ports.

The return value indicates the result of the data transfer. It can be one of the following values:

ERROR_OK

= 0

successful data transfer

ERROR_INTERFACE

= 1

communication port is not initialized

ERROR_COMMUNICATION

= 3

faulty data transfer

ERROR_INVALID_PARAM

= 4

invalid parameter