MCA166-USB

Command name

CMD_SET_REPEAT

Description

The command sets the number(rep: 0 ... 65535) of sweeps for repetitive
measurement.

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

rep

0

end flag

Byte String (HEX)

A5

5A

4A

00

rep

l

rep

h

00

00

00

00

B9

9B

Used setup value

None

Affected setup value

Repeat

MCA state

STATE_READY

Command name

CMD_SET_MCS_CHANNEL

Description

The command set the number of channels(ch: 128 ... 4096) for MCS mode

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

ch

0

end flag

Byte String (HEX)

A5

5A

63

00

ch

l

ch

h

00

00

00

00

B9

9B

Used setup value

None

Affected setup value

mcs_channels

MCA state

STATE_READY

Command name

CMD_SET_TIME_PER_CHANNEL

Description

The command sets the dwell time(tpc: 1 ... 65535) per channel for the MCS

mode

40

.

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

tpc

0

end flag

Byte String (HEX)

A5

5A

4B

00

tpc

l

tpc

h

00

00

00

00

B9

9B

Used setup value

None

Affected setup value

time_per_channel

MCA state

STATE_READY

40 It is the real time per channel multiplied with 10