GBS Elektronik MCA166-USB User Manual

Page 51

Advertising
background image

Name

MMCA_QUERY_POWER

Description

The command reads the MMCA power state.

Parameter

-

Result array

Battery current [mA]

long value

Byte Offset 4

HV primary current [mA]

long value

Byte Offset 8

+12 primary current [mA]

long value

Byte Offset 12

-12 primary current [mA]

long value

Byte Offset 16

+24 primary current [mA]

long value

Byte Offset 20

-24 primary current [mA]

long value

Byte Offset 24

Battery voltage [mV]

long value

Byte Offset 28

HV [V]

27

long value

Byte Offset 32

HV state

long value

Byte Offset 36

Byte Offset 40

Current high voltage [V]

Byte Offset 44

Byte Offset 48

Power Switches

long value

0X80 -24V

ON

0X40 +24V

ON

0X20 -12V

ON

0X10 +12V

ON

0X02

HV

ON

Byte Offset 52

Charger current [mA]

long value

...

Byte Offset 64

Battery current [mA] at stop

long value

Byte Offset 68

HV primary current [mA] at stop

long value

Byte Offset 72

+12 primary current [mA] at stop

long value

Byte Offset 76

-12 primary current [mA] at stop

long value

Byte Offset 80

+24 primary current [mA] at stop

long value

Byte Offset 84

-24 primary current [mA] at stop

long value

Byte Offset 88

Battery voltage [mV] at stop

long value

Byte Offset 92

HV [V]

28

at stop

long value

Byte Offset 116

Charger current [mA] at stop

long value

...

Byte Offset 128

mca_state

29

integer value

Return Code

see MMCA_RESET command

27 multiply value with 1.2 V
28 multiply value with 1.2 V

29 see MMCA_QUERY_SYSTEM_DATA

51