GBS Elektronik MCA166-USB User Manual

Page 64

Advertising
background image

MCA166-USB

Command name

CMD_SET_MCS_INPUT

Description

The command sets the input for the Multi-Channel Scaler.
ip = 0 external TTL signals

ip = 1 internal count rate signal
ip = 2 counts only pulses with amplitude between LLD and ULD threshold

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

ip

0

end flag

Byte String (HEX)

A5

5A

55

00

ip

00

00

00

00

00

B9

9B

Used setup value

None

Affected setup value

mca_input_mcs

MCA state

STATE_READY

Command name

CMD_SET_THRESHOLD

Description

This command set the analogue threshold value (thr: 2 …60).

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

thr

0

end flag

Byte String (HEX)

A5

5A

47

00

thr

00

00

00

00

00

B9

9B

Used setup value

None

Affected setup value

Threshold

MCA state

STATE_READY

Command name

CMD_SET_SHAPING_TIME

Description

The command sets the amplifier shaping time(dtc).

dtc = 1 shaping time low (1 µsec)
dtc = 3 shaping time high (2 or 3 µsec)

41

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

dtc

0

end flag

Byte String (HEX)

A5

5A

52

00

dtc

00

00

00

00

00

B9

9B

Used setup value

None

Affected setup value

Dtc

MCA state

STATE_READY

41 The values may be different for some MCA166-USB configurations.