MCA166-USB

MCA Stabilisation Commands:

Name

MMCA_SET_STABILISATION

Description

The command sets the peak stabilisation.

1. Parameter

Stabilisation flag

Integer
value

0 turns stabilisation off
1 turns stabilisation on (keep peak at current

position)
>1 stabilisation channel (move peak to the

stabilisation

channel and hold peak at the

stabilisation channel)

2. Parameter

Peak ROI begin

Integer

value

LLD<= begin < end

3. Parameter

Peak ROI end

Integer
value

begin < end <= ULD, (end-begin) < 250

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_SET_STAB_PARAM

Description

The command sets the peak stabilisation parameters.

1. Parameter

Stabilisation time

Integer

value

default 10 sec interval

2. Parameter

Stabilisation area

long value

default 25000

Results

-

Return Code

see MMCA_RESET command

MCA Detector Power Commands:

Name

MMCA_SET_PREAMPLIFIER_POWER

Description

The command sets the preamplifier power.

1. Parameter

Preamplifier switch integer value

0X80 -24V
0X40 +24V

0X20 -12V
0X10 +12V

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_SET_HIGH_VOLTAGES

Description

The command sets the detector high voltage and controls the HV-inhibit-signal.

20

1. Parameter

High voltage

integer value

0 ... 3000

2. Parameter

High voltage inhibit long value

i = 0 Inhibit off

i = 1 ”Canberra-/ DGF-mode”, HV shut
down

if inhibit signal high

i =-1 ”Ortec-mode”, HV shut down

if inhibit signal low

Results

-

20 The high voltages will ramp up or down by the MMCA.

46