MCA166-USB

Command name

CMD_SET_PZC

Description

The command performs the Pole Zero Cancellation.
pv: 0 ... 2499 PZC control voltage

pm = 0

set PCZ only

pm > 0

set PZC and measure PZC offset

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

pm

Pv

end flag

Byte String (HEX)

A5

5A

58

00

pm

l

pm

h

pv

l

pv

h

00

00

B9

9B

Used setup value

None

Affected setup value

pzc_value

Results

If pm > 0 the function returns with the number of counts and the - PZC offset.

MCA state

STATE_READY

Command name

CMD_SET_PZC_TIME_OFFSET

Description

The command sets the time offset for Pole Zero Cancellation.

t1: 0 ... 31

Time offset for low shaping time

t2: 0 ... 31

Time offset for high shaping time

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

t1

t2

end flag

Byte String (HEX)

A5

5A

60

00

t1

00

t2

00

00

00

B9

9B

Used setup value

None

Affected setup value

pzc_dtc_1_offset, pzc_dtc_3_offset

MCA state

STATE_READY