MCA166-USB

Strip:

Strip subtracts a specified fraction of the spectrum. The Strip factor can be “positive“ or
“negative“ :

Spectrum = Spectrum - F*Spectrum

y

'

y

x

where

Spectrum

y

'

- Count content of channel i in the resulting spectrum

Spectrum

y

- Count content of channel i in the original spectrum

F - Strip factor

Spectrum

x

- Count content of channel i in the spectrum to be subtracted

E-calibration:

The Energy calibration allows to convert the x-axis channel numbers into energy values
in KeV. The following formula is used:

E = S Channel + O

where,
S - Slope
O - Offset

The coefficients can be set as numeric values or by linear fit of two energy
channel/peak centroid pairs.

34