CMG-3ESPCD Digital Broadband Seismometer

Using Scream!

The baud rate you choose must satisfy two conditions: •

It must be high enough to allow all the transmission of all data
generated by the digitiser at the sampling rates you have chosen. For

three streams of data at 100 Hz, for example, 9,600 baud will usually be
sufficient. If you wish to transmit 200 Hz data, however, the baud rate

must be at least 19,200.

It must be low enough to fit within the operating range of the telemetry

equipment you are using. While modern modems often offer transfer
rates up to 56 kBaud, the telephone or transmission lines may not

support these rates. The same holds true for radio telemetry.

Usually, the transmit and receive rates of the data port will be the same. If

not, you may select different data rates by removing the tick in the check-box
marked Identical TX/RX rates.

The Stop Bits option allows you to choose whether the serial link uses 1 or 2
stop bits. In most cases this can be left at 1, although 2 may be required if you

are sending data over ‘difficult’ transmission lines (for example, some types of
radio link). Using 2 stop bits will add a 10% overhead to the data.

You will also need to set the data rate for Scream's local serial port, as well as
for the EAM or other communications device (if you are using one). In

Scream!, you can configure a serial port by right-clicking on its icon (not that
of the digitiser) and selecting Configure... from the pop-up menu: for more

details, consult the on-line help or user guide for Scream!. If you are using an
additional communications device, you should consult its documentation to

learn how to set its baud rate.