Command line interface

will calculate short-term averages for 1 second of the Z and N components,
and 2 seconds of the E and auxiliary components.

6.4.4 LTA

Syntax: Z­secs N­secs E­secs X­secs LTA

Sets the length of the “long-term” averaging period in the STA/LTA triggering
algorithm.

The arguments Z­secs, N­secs, E­secs and X­secs are the time period
over which to calculate the average for the Z, N, E, and auxiliary components

respectively. If a component is not considered by the triggering algorithm (see
TRIGGERS, above), the value you specify here will be ignored. For example,

15 20 20 20 STA

will calculate long-term averages for 15 seconds of the Z component and 20

seconds of the N, E and auxiliary components.

6.4.5 RATIOS

Syntax: Z­ratio N­ratio E­ratio X­ratio RATIOS

Sets the ratio of STA to LTA above which a trigger will be declared in the

STA/LTA triggering algorithm.

The arguments Z­ratio, N­ratio, E­ratio and X­ratio are the time

period over which to calculate the average for the Z, N, E, and auxiliary
components respectively. If a component is not considered by the triggering

algorithm (see TRIGGERS, above), the value you specify here will be ignored.
For example,

4 10 10 10 RATIOS

will cause the digitiser to trigger if the STA/LTA ratio is above 4 for the Z

component or above 10 for the remaining components.

6.4.6 PRE-TRIG

Syntax: time PRE­TRIG

Sets the pre-trigger recording time. The argument time is the number of

seconds of data to output from before a trigger is declared.

6.4.7 POST-TRIG

Syntax: time POST­TRIG

Sets the post-trigger recording time. The argument time is the number of

seconds of data to output after a trigger condition lapses. If an event persists

MAN-C3E-0002