Guralp Systems CMG-3ESPCD User Manual

Page 73

Advertising
background image

CMG-3ESPCD Digital Broadband Seismometer

Command line interface

first; the digitiser will only transmit the next block when it receives an
acknowledgement of the previous block.

6.7.6 DOWNLOAD

Syntax: DOWNLOAD

Sets up a data transfer from the Flash memory over the serial connection.
Which data are downloaded depends on various parameters you can set,

which allow you to select a particular stream, streams of a specified sample
rate, or streams within a certain time window. You can set parameters

separately, or string the definitions before the DOWNLOAD command, e.g.

ALL-FLASH STREAM HPA0N1 DOWNLOAD

2004 12 01 00 00 FROM-TIME ALL-DATA DOWNLOAD

100 S/S ALL-TIMES DOWNLOAD

DOWNLOAD with no arguments starts another download with the same
parameters as last time. See below for full details of the parameter

commands.

The DOWNLOAD command returns immediately, so that you can issue more

commands if required. To close the connection and begin downloading, issue
the GO command. You can pause the download by entering terminal mode,

and restart with another GO or abort with END­DOWNLOAD.

6.7.7 ALL-FLASH

Syntax: ALL­FLASH

Sets the read point to the oldest data held by the digitiser. This command

does not alter which streams are to be transmitted; you should specify streams
or use the ALL­DATA command in addition to this one.

6.7.8 ALL-DATA

Syntax: ALL­DATA

Instructs the digitiser to transmit all the data streams it holds next time a
DOWNLOAD is issued. This command does not alter the read point or time

period; you should specify a time period or use the ALL­FLASH command in
addition to this one.

6.7.9 STREAM

Syntax: STREAM stream­id (n.b.)

Instructs the digitiser to transmit only the stream with ID stream­id.
Stream IDs are normally a 4-character device code (e.g. HPA0) followed by a

MAN-C3E-0002

73

Issue B - November 2013

Advertising
This manual is related to the following products: