1 sensor response – Guralp Systems CMG-3ESPCD User Manual

Page 7

Advertising
background image

CMG-3ESPCD Digital Broadband Seismometer

Introduction

port or to a Güralp Systems data module using the cable supplied. The
digitiser also controls the instrument's centring and mass locking processes.

Each seismometer is delivered with a detailed calibration sheet showing its
serial number, measured frequency response in both the long period and the

short period sections of the seismic spectrum, sensor DC calibration levels,
and the transfer function in poles/zeros notation.

2.1

Sensor response

The 3ESPCD can be supplied with a response which is flat to velocity from

100 Hz to any of 0.1 Hz (10 s), 0.033 Hz (30 s), 0.016 Hz (60 s), 0.01 Hz (100 s)
or 0.0083 Hz (120 s). Alternatively, a hybrid response function may be

provided. See section 7.1.1 on page 77 for more details.

If you do not require high-frequency data, a low-pass filter may be installed at

a frequency (below 100 Hz) that you specify.

MAN-C3E-0002

7