CMG-3ESPCD Digital Broadband Seismometer

Command line interface

Because of the way FORTH works, you can insert MINUTE commands into
SQUAREWAVE, SINEWAVE or RANDOMCAL commands, for example:

N/S 4 HZ 5 MINUTE SINEWAVE E/W 10 MINUTE SQUAREWAVE

6.5.5 %AMPLITUDE

Syntax: value %AMPLITUDE

Sets the relative amplitude of the signal, as a percentage. For SQUAREWAVE

and RANDOMCAL calibrations, the default of 100% is normally suitable. If you
have a high-gain sensor, you may have to use a lower value. For SINEWAVE

calibrations, you may have to adjust the %AMPLITUDE setting to suit the
frequency of the calibration signal.

You can insert %AMPLITUDE commands into calibration commands as
described above, e.g.:

Z 3 MINUTE 50 %AMPLITUDE RANDOMCAL

6.6

Actions

The commands in this section trigger specific actions, such as mass-locking.

6.6.1 LOCK

Syntax: LOCK

Locks the sensor mass of the instrument. Whilst this process is executing, the

system displays the mass position for each component, measuring one
position each second (i.e. one reading every 4 seconds for each component of

a 4-component instrument). The process returns when locking is complete.
The mass positions are measured to 16-bit accuracy, with full deflection

corresponding to values around ±32000 counts.

6.6.2 UNLOCK

Syntax: UNLOCK

Unlocks the sensor mass, and returns when unlocking is complete. The

system will display the mass positions as the process continues, as for LOCK,
above.

6.6.3 CENTRE

Syntax: CENTRE

Centres the sensor mass, and returns when centring is complete. The system
will display the mass positions as the process continues, as for LOCK, above.

When the mass is correctly centred, the readings should be less than ±1000
counts.

MAN-C3E-0002

70

Issue B - November 2013