5. Operating Instructions

5.1 Introduction to the operation

Programs can be selected with P1, P2, P3,
and P4 keys. The respective LED will then be
illuminated.
The parameters are shown in the display. The
program is started by pressing Start. The
display then indicates the remaining time of
the program. The program can be inter-
rupted by pressing Stop.

5.2 Working with Program P1 (set

program with plaster model)

Tempering program for SR Adoro materials
(light/heat), which are polymerized/tempered
on a plaster model (the parameters are set by
the manufacturer and cannot be altered).

If the restorations are placed in the curing
chamber together with a plaster model,
Program P1 must be used (see also sticker on
the apparatus).

When tempering partial models
(e.g. 1/4 models, single or multiple
cores), please refer to the additio-
nal instructions in the SR Adoro
Instructions for Use.

5.3 Working with Program P2

(set program)

Light polymerization program for light-curing
resins (light). (The parameters are set by the
manufacturer and cannot be altered).

5.4 Working with Program P3 (set

program without plaster model)

Tempering program for SR Adoro materials
(light/heat), which are polymerized/tempered
without a plaster model (the parameters are
set by the manufacturer and cannot be
altered).

If the restorations are placed in the curing
chamber without the plaster model, Program
P3 must be used (see also sticker on the
apparatus).

5.5 Working with Program P4

(individual program)

The parameters for Program P4 can be set as
desired (see also 6.2.2).

Program Table

No

1

2

3

4

Program

SR Adoro tempering
program (with plaster
model)

Light-curing
program

SR Adoro tempering
program (without
plaster model)

Individual curing
program

VB min

10:00

––––

10:00

0:00 to
10:00***

HP

set
****

––––

set
****

set
****

BP min

––––

10:00

––––

0:00 to
30:00**

VG min

7:00

––––

7:00

0:00 to
30:00***

Stage

3

––––

3

0 – 3

AP min

5:00
****

1:00
****

5:00
****

****
1:00/5:00*

VB

HP

VG

AP

BP

Stage 0

Stage 1

Stage 2

Stage 3

* depending on the heating stage;

** only at stage 0,

*** only at stage 1, 2, 3,

**** cannot be set

VB = Precuring,

HP = Heating,

VG = Tempering,

AP = Cooling,

BP = Curing

Heat performance:*
Stage 0 = Heater off
Stage 1 = approx. 80 °C / 176 °F
Stage 2 = approx. 95 °C / 203 °F
Stage 3 = approx. 104 °C / 219 °F

* The temperatures were measured with a reference tooth

(typical value) with a plaster model (P1) and without
plaster model (P3).