7. Maintenance, Cleaning, and Diagnosis

Item

Frequency

Cleaning material/measure

Dust filter

Monthly

Batting or replacing

Housing surfaces

If required

Cloth

Magnet and counterpiece

If required

Cloth

Reflector

If required

Cloth

7.2 Cleaning

The unit may only be
cleaned when it is
cool, since there is
burn hazard. Do not

use any cleaning solutions. The
heater must not be touched, as
hand sweat considerably
shortens the life cycle of the
unit.

What

Part

Frequency

Check all plug-in connections for correct fit.

Power plug

Weekly

Check the switch-off mechanism by opening the curing chamber during

Interlock

Weekly

program operation. If the fan continues to work, the safety switch
(interlock) is defective. In this case, contact your local Ivoclar Vivadent
Service Center.

Check the infrared heater (26) for damage of the glass.

Infrared heater (26)

Before the first
operation. Weekly

Check temperature and light performance

Test Set

Every 6 months

This chapter describes the user maintenance
and cleaning procedures. All other tasks
must be performed by qualified service
personnel at a certified Ivoclar Vivadent
Service Center.

7.1 Monitoring and maintenance

The time for these maintenance procedures
depends on the frequency of use and the
working habits of the users. For that
reason, the recommended times are only
approximates.

Disconnect power before maintenance
and cleaning, since there is a risk of
electrical shock.

The following parts have to be cleaned from time to time:

Use only original Ivoclar Vivadent dust