Dual electronic expansion valve module, Confi guration screens, Technical guide – Orion System Dual Electronic Expansion Valve Module User Manual

Page 17

Advertising
background image

Technical Guide

Dual Electronic Expansion Valve Module

17

Confi guration Screens

Confi guration Screens

Refer to the following map when navigating through the Confi guration
Screens. From the CONFIG Screen, press

<ENTER> to scroll through

the screens and change setpoints. Use the

<UP> and <DOWN> arrow

keys to change your selections. Press

<ENTER> to save any changes.

CONFIG

ADDRESS

1 TO 4

VALVE B

ENABLED/

DISABLED

VLVSTEPS

1596,2500,

3193,6386

MAX VLV%
0 TO 100

MIN VLV%
0 TO 100

SP-A CAL

-10 TO 10

MAXIMUM VALVE POSITION

The maximum position each valve will modulate (%).

Default =100

MINIMUM VALVE POSITION

The maximum position each valve will modulate (%).

Default = 0

SUCTION PRESSURE VALVE A

CALIBRATION OFFSET

If the Suction Pressure Sensor is reading incorrectly,

you can use this offset to adjust its reading.

Default = 0

SP-B CAL

-10 TO 10

SUCTION PRESSURE VALVE B

CALIBRATION OFFSET

If the Suction Pressure Sensor is reading incorrectly,

you can use this offset to adjust its reading.

Default = 0

VALVE STEPS

Confi gurable for what valve is being used

(1596, 2500, 3193, 6386)

Default = 2500

CURRENT ADDRESS OF THE BOARD

The address confi guration is not used in Stand

Alone Mode. Stand Alone Mode Default is 1.

VALVE B ENABLED/DISABLED

When using the module for only one valve, valve
B can be disabled so false information is not dis-

played such as alarms and sensor readings.

Advertising