Ac filter – RIGOL M300 User Manual
Page 52

RIGOL
© 2007 RIGOL Technologies, Inc.
User’s Guide for DM3000 Series
2-4
AC Filter
There are three settings at the AC Filter menu. Choose the appropriate setting for
more accurate measurements. This function applies AC Voltage and AC Current
measurement only.
Press
Filter, enter the menu shown below:
Figure 2- 2 Set the AC Filter
Table 2- 2 Menu Description
Function
Menu
Description
Slow
Set up the filter with low speed.
Mid
Set up the filter with to middle speed.
Fast
Set up the filter with high speed.
Save all changes, back to a higher level menu.
Table 2- 3 AC Filter Parameters Characteristics
AC Filter Options
Input Frequency
Setting Timer
Slow
3Hz~300kHz
1.2 reading/s
Mid
20Hz~300kHz
0.5 reading/s
Fast
200Hz~300kHz
0.3 reading/s
AC Filter
The AC Filter Parameters are stored in nonvolatile memory. The default is “Mid”
(middle).