4 calibrate<chn>:date:due? [min|max|def] 100, Calibrate<chn>:date:due? [min|max|def – Fluke 1529 Chub-E4 User Manual

Page 106

Advertising
background image

2000,1,1. Specifying the MAX parameter returns a value of 2099,12,31. Spec-
ifying the DEF parameter returns a value of 2000,1,1.

8.4.5.3

CALibrate<chn>:DATE:CALibrate
(<year>,<month>,<day>)|MIN|MAX|DEF

Sets the calibration date for the specified channel.

CAL1:DATE:CAL 2000,8,29

The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. The <year>, <month>, and <day> parame-
ters must all be supplied. The <year> parameter is a four-digit number (2000 to
2099). The <month> parameter is a one or two-digit number (1 to 12). The
<day> parameter is a one or two-digit number (1 to 31). Specifying the MIN,
MAX or DEF parameter sets the calibration date to the minimum, maximum,
or default value respectively.

This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN <pass> command
(see Section 8.4.11, Password Commands).

8.4.5.4

CALibrate<chn>:DATE:DUE? [MIN|MAX|DEF]

Returns the recalibration date for the specified channel.

CAL1:DATE:DUE?

2001,9,21

The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. The response is returned in the format,
<year>,<month>,<day>. Specifying the MIN parameter returns a value of
2000,1,1. Specifying the MAX parameter returns a value of 2099,12,31. Spec-
ifying the DEF parameter returns a value of 2000,1,1.

8.4.5.5

CALibrate<chn>:DATE:DUE (<year>,<month>,<day>)|MIN|MAX|DEF

Sets the recalibration date for the specified channel.

CAL1:DATE:DUE 2001,8,28

The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. The <year>, <month>, and <day> parame-
ters must all be supplied. The <year> parameter is a four-digit number (2000 to
2099). The <month> parameter is a one or two-digit number (1 to 12). The
<day> parameter is a one or two-digit number (1 to 31). Specifying the MIN,
MAX or DEF parameter sets the recalibration date to the minimum, maximum,
or default value respectively.

This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN <pass> command
(see Section 9.4.11, Password Commands).

1529 Chub-E4 Thermometer Readout