6 sense<chn>:resistance:range <num>|min|max|def 92, 7 sense<chn>:resistance:wire? [min|max|def] 92, Sense<chn>:resistance:range <num>|min|max|def – Fluke 1529 Chub-E4 User Manual

Page 98: Sense<chn>:resistance:wire? [min|max|def

Advertising
background image

0

This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.

The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5
to 14). Specifying the MIN parameter returns a value of 0. Specifying the MAX
parameter returns a value of 1. Specifying the DEF parameter returns a value of
0. The *RST command sets the resistance range to the default (DEF).

8.4.2.6

SENSe<chn>:RESistance:RANGe <num>|MIN|MAX|DEF

Sets the resistance range setting.

SENS4:RES:RANG MIN

This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.

The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5
to 14). The <num> parameter must be a number between 0 and 1. Specifying
the MIN, MAX or DEF parameter sets the resistance range setting to the mini-
mum, maximum, or default value respectively.

This command may be password-protected. To access this command:

• If the SYST:PASS:CONV? command returns a 0 (probe password-protec-

tion is disabled), the password is not required.

• If the SYST:PASS:CONV? command returns a 1 (probe password-protec-

tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN <pass> command (see Section 9.4.11, Password Com-
mands).

8.4.2.7

SENSe<chn>:RESistance:WIRE? [MIN|MAX|DEF]

Returns the probe wire setting.

SENS1:RES:WIRE?

3

This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.

The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5
to 14). Specifying the MIN parameter returns a value of 2. Specifying the
MAX parameter returns a value of 4. Specifying the DEF parameter returns a
value of 4.

1529 Chub-E4 Thermometer Readout
User’s Guide

92