5 system reset 77, Figure 48 pass cal reset 77, Reset 77 – Fluke 1529 Chub-E4 User Manual

Page 83: System reset 77, System reset, Figure 48, Pass cal reset, 159c, 142c, 634c

Advertising
background image

The C0K, C10K, and C100K parameters apply to the thermistor probe
types and 10K

Ω resistance range. Use these parameters to adjust the mea-

surement at 0K,10K, and 100K

Ω respectively. These values are in Kohms.

The DATE parameter is the date the current calibration was performed.

The DUE parameter is the date the current calibration expires.

For thermocouple channels, the parameters C0, C100, CRJ, DATE, and DUE
are displayed.

The C0 and C100 parameters adjust the measurement at 0 and 100 mV re-
spectively. These values are in mV.

The CRJ parameter adjusts the reference junction temperature measure-
ment. This parameter is in °C.

The DATE parameter is the date the current calibration was performed.

The DUE parameter is the date the current calibration expires.

Press the

LR buttons to change a setting and the UD buttons to scroll be-

tween parameters. Press the Enter button to save the new setting. Press the
Exit button to exit to the menu.

7.5.5

System Reset

The SYSTEM RESET function resets all operating parameters to the factory
default values. This function is password-protected (see Section 7.5.3, Pass-
word). In order to access the system reset, you must enter the correct four-digit
password. When this function is selected, the bottom portion of the display
prompts the user to enter the current password to gain access to this option.

77

7 Menu Functions

System Menu

1

24.159C

3

24.142C

2

24.634C

4

24.015C

ENTER CORRECT PASSWORD FOR ACCESS.

PASSWRD:

0

EXIT

ENTER / MENU

UNIT

CONTRAST

1529

CHUB E-4 THERMOMETER READOUT

Figure 48 Pass Cal Reset